Z-0066-2019: OptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Ser

Recall record Z-0066-2019 for Johnson & Johnson Surgical Vision Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0066-2019
Company in FDA record
Johnson & Johnson Surgical Vision Inc
FDA record creation date
Oct 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.

Source facts

Action
On May 9, 2018, affected customers in the U.S. and Puerto Rico were contacted by phone and verbally informed to stop using the affected system. On May 22, 2018, the firm, Johnson & Johnson Vision, sent an "URGENT MEDICAL DEVICE FIELD REMOVAL" letter dated May 18, 2018 via Federal Express overnight delivery with signature required to all customers with a system containing the affected software installed. The customers were instructed to: STOP Using the affected system and complete and return the Customer Reply Form via fax to: Johnson & Johnson Vision Quality Assurance at 1-714-247-4510, emailed to RegComplione@its.jnj.com, or provide to your Johnson & Johnson Vision Representative within 3 business days of receipt of the letter. The firm will coordinate the removal of the affected software for all affected systems. All impacted customers will be provided with the new/updated software version. If you have any questions, contact the Director, Quality Assurance at 408-273-4100 or email: creyes51@its.jnj.com.
Initiated
2018-05-09
Root cause
Device Design
Terminated
2021-01-13
Distribution
US Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany.
Recall number
Z-0066-2019
Recall status
Terminated
CDRH record ID
166720
FDA record created
2018-10-03
Recall event number
80849

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗