Open original FDA record ↗Recall record
- FDA identifier
- Z-0066-2019
- Company in FDA record
- Johnson & Johnson Surgical Vision Inc
- FDA record creation date
- Oct 3, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.
Source facts
- Action
- On May 9, 2018, affected customers in the U.S. and Puerto Rico were contacted by phone and verbally informed to stop using the affected system. On May 22, 2018, the firm, Johnson & Johnson Vision, sent an "URGENT MEDICAL DEVICE FIELD REMOVAL" letter dated May 18, 2018 via Federal Express overnight delivery with signature required to all customers with a system containing the affected software installed. The customers were instructed to: STOP Using the affected system and complete and return the Customer Reply Form via fax to: Johnson & Johnson Vision Quality Assurance at 1-714-247-4510, emailed to RegComplione@its.jnj.com, or provide to your Johnson & Johnson Vision Representative within 3 business days of receipt of the letter. The firm will coordinate the removal of the affected software for all affected systems. All impacted customers will be provided with the new/updated software version. If you have any questions, contact the Director, Quality Assurance at 408-273-4100 or email: creyes51@its.jnj.com.
- Initiated
- 2018-05-09
- Root cause
- Device Design
- Terminated
- 2021-01-13
- Distribution
- US Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany.
- Recall number
- Z-0066-2019
- Recall status
- Terminated
- CDRH record ID
- 166720
- FDA record created
- 2018-10-03
- Recall event number
- 80849