Issues with the Ingenuity TF PET/CT Software Version 4.0 could pose a risk for patients in the event that a patient rescan were to require reinjection of PET radiopharmaceutical.
Source facts
Action
Philips Healthcare sent an Field Safety Notice dated August 16. 3013, via certified mail to all US customers. Market Representative will distribute the letter internationally. The letter identified the product the problem and the action needed to be taken by the customer.
Field Safety Notice 88200460 is intended to inform you about;
" what the problem is and under what circumstances it can occur
" the actions that should be taken by the customer / user in order to prevent risks for patients or users
" the actions planned by Philips to correct the problem
Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
If you need any further information or support concerning these issues, please contact your local Philips representative or local Philips Healthcare office.
For North America and Canada contact the Customer Care Solutions Center
(1-800.722.9377, option 5: Enter Site ID or follow the prompts).
Initiated
2013-08-16
Root cause
Software design
Terminated
2014-03-26
Distribution
Worldwide Distribution - US (nationwide) including the states of NY, OH and Pa, and the countries of Australia, Austria, Egypt, France, Germany, India and Netherlands.
Recall number
Z-0067-2014
Recall status
Terminated
CDRH record ID
121747
FDA classified
2013-10-25
Recall event number
66205