Open original FDA record ↗Recall record
- FDA identifier
- Z-0067-2017
- Company in FDA record
- Brathwaites Oliver Medical
- FDA classification date
- Oct 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Introes Pocket Bougie an endotracheal tube introducer size 14 Fr (4.7mm) x 60 cm may break during intubation.
Source facts
- Action
- On August 19, 2016 the firm sent a MEDICAL DEVICE RECALL NOTICE & CORRECTIVE ACTION letter to affected customers. The letter described the reason for the recall and necessary actions to be taken by the consignees. These included: BOMImed is requesting bougies with lot# 101615 to be returned for evaluation due a potential manufacturing concern. Complete the included Medical Device Recall Customer Letter and follow the instructions to obtain a Return Material Authorization number and applicable shipping instructions to return the affected products. BOMImed Regulatory department also requires the below Customer Confirmation section to be completed and return to regulatory@bomimed.com to confirm this notification was received.
- Initiated
- 2016-08-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-07-19
- Distribution
- U.S. Consignee located in the following states: IL, WI, WE, OR, NY, UT, IN, NE, PA, MO, OH, AZ, MA, CT and NC.*** Foreign Consignee: Canada.
- Recall number
- Z-0067-2017
- Recall status
- Terminated
- CDRH record ID
- 149464
- FDA classified
- 2016-10-11
- Recall event number
- 75148