Due to manufacturing issues with the liquid skin adhesive containing in surgical convenience trays/kits.
Source facts
Action
On 01/15/2026, a customer notification letter was email to a single customer informing them that, the MASTISOL Liquid adhesive shipped in kits in lot 1225271, P.O.#2513972 was exposed to a double sterilization cycle by Ferndale Laboratories, Inc. which may contribute to the butyrate tubing being brittle and cracking during actuation.
The customer was instructed to:
1. lmmediately examine your inventory and quarantine Lot 1225271.
2. Return all units of MASTISOL in Lot 1225271 to:
CIPM
12460 Network Blvd. Suite 106
San Antonio, TX 78249
3. Complete the enclosed Return Materials Authorization Form QF-85-05-02 and return to denise@cipmtrays.com or quality@cipmtrays.com.
For questions - call Toll Free 1-888-392-4514
Initiated
2026-01-15
Root cause
Process control
Distribution
U.S. Nationwide distribution in the state of TX.
Recall number
Z-0067-2027
Recall status
Open, Classified
CDRH record ID
222233
FDA classified
2026-10-01
Recall event number
99699
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.