The image balance center point correction algorithm used in CT image reconstruction may cause a disapperance or alteration of small circular anatomical structures that are both near the image center and perpendicular to the image plane. This artifact may lead to the misinterpretation of anatomy at the center of the image.
Source facts
Action
The recalling firm sent a recall/field correction letter, dated August 18, 2005 to customers, notifying them of the problem.
Initiated
2005-08-18
Root cause
Other
Terminated
2008-10-29
Distribution
The product was distributed to customers within the US and to foreign customers located in Canada, France, Thailand, South Africa, and Italy.
Recall number
Z-0068-06
Recall status
Terminated
CDRH record ID
41582
FDA classified
2005-10-21
Recall event number
33273