Open original FDA record ↗Recall record
- FDA identifier
- Z-0068-2018
- Company in FDA record
- Limacorporate S.p.A
- FDA record creation date
- Nov 6, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An internal analysis on complaints received indicated that the welding process may have been performed sub optimally in the 12 lot # involved , leading to reduced mechanical strength of the device and possible breakage during impaction when the instrument is subjected to repeated multi-axial stresses.
Source facts
- Action
- The Firm, Lima Corporate has notified the US agents of the recall and asked to check each single set with the affected product codes, identify the items and remove them from the inventory. The US distributors have also been instructed to fill out and sign a response document specifying the quantities returned. For further questions, please call (800) 962-2578.
- Initiated
- 2017-09-08
- Root cause
- Vendor change control
- Distribution
- Worldwide Distribution - US Distribution and to the countries of : Australia, Austria, Belgium, Brazil, Czech Republic, Germany, Italy, New Zealand, Poland, Russia, Slovakia, Slovenia, Spain, South Africa, Switzerland, Turkey, UK and Germany
- Recall number
- Z-0068-2018
- Recall status
- Completed
- CDRH record ID
- 158680
- FDA record created
- 2017-11-06
- Recall event number
- 78109