Open original FDA record ↗Recall record
- FDA identifier
- Z-0068-2026
- Company in FDA record
- GE Medical Systems, SCS
- FDA classification date
- Oct 8, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity, and availability of patient data.
Source facts
- Action
- On September 5, 2025, firm began deployment of customer letter. Customers were informed they could continue use of their device by following the instructions provided: 1. Ensure the AW Server is installed on a facility internal network, that is not accessible from any public network. If access to AW Server from outside the facility network is desired, this connectivity must be configured to be usable exclusively through the facility s Virtual Private Network (VPN) or equivalent network security solution as described in your AW Server s Pre-Installation Manual that is available in the Customer documentation portal. NOTE: Access management to the facility s internal network on which the AW Server is installed on, and management of the network security solution is under the strict responsibility of the facility. 2. Ensure strong, secure passwords are configured for the AW Server. 3. GE HealthCare recommends that organizations follow security and cybersecurity best practices, including securing the network where AW Server is connected.
- Initiated
- 2025-09-05
- Root cause
- Software in the Use Environment
- Distribution
- Domestic US distribution nationwide. International distribution worldwide.
- Recall number
- Z-0068-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216255
- FDA classified
- 2025-10-08
- Recall event number
- 97657