Skip to content

Public FDA record

Z-0068-2027: NovaSeq 6000Dx including DRAGEN Server, Catalog Number: 20068232; Illumina DRAGEN Server for NextSeq 55

Recall record Z-0068-2027 for Illumina, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0068-2027
Company in FDA record
Illumina, Inc.
FDA classification date
Oct 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro diagnostic assays has a server with a security vulnerability.

Source facts

Action
On 9/17/2024, and 1/16/2025 field safety notices were emailed to customers who were asked to take actions on webpage: https://support.illumina.com/support-content/bmc.html Customers were also asked to complete and return the Verification Form via email to techsupport@illumina.com This email can also be used to ask questions or to report potentially compromised servers.
Initiated
2024-09-17
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of TX, KS, NC, CA, UT, KA, MA, NJ, MD and the countries of Spain, South Korea, Japan, Australia, Italy, United Arab Emirates, Sweden, United Kingdom, Czech Republic, Singapore, Austria, Poland, France, Germany, Finland, Switzerland, Slovakia.
Recall number
Z-0068-2027
Recall status
Open, Classified
CDRH record ID
221916
FDA classified
2026-10-01
Recall event number
99585

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗