- Recall classification
- Class II
Reason stated in the source
Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
Source facts
- Action
- On 08/20/18, the firm, icumedical, mailed "URGENT MEDICAL DEVICE RECALL" l notices to consignees. The notice described the product, problem and actions to be taken. If product has been further distributed, the firm requested consignees to notify users. Customers were advised to return affected product only if the blood filter is needed. Customers needing response forms and/or return labels were advised to call 1-800-479-0551. For further inquiries call: 1-866-829-9025.
- Initiated
- 2018-08-20
- Root cause
- Device Design
- Distribution
- U.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
- Recall number
- Z-0069-2019
- Recall status
- Completed
- CDRH record ID
- 167075
- FDA record created
- 2018-10-04
- Recall event number
- 80946