Z-0069-2019: ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascul

Recall record Z-0069-2019 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0069-2019
Company in FDA record
ICU Medical, Inc.
FDA record creation date
Oct 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.

Source facts

Action
On 08/20/18, the firm, icumedical, mailed "URGENT MEDICAL DEVICE RECALL" l notices to consignees. The notice described the product, problem and actions to be taken. If product has been further distributed, the firm requested consignees to notify users. Customers were advised to return affected product only if the blood filter is needed. Customers needing response forms and/or return labels were advised to call 1-800-479-0551. For further inquiries call: 1-866-829-9025.
Initiated
2018-08-20
Root cause
Device Design
Distribution
U.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
Recall number
Z-0069-2019
Recall status
Completed
CDRH record ID
167075
FDA record created
2018-10-04
Recall event number
80946

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗