Z-0069-2021: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precice Bone Transport Antegrade Femur Trochanter 10 Degree Bend

Recall record Z-0069-2021 for Nuvasive Specialized Orthopedics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0069-2021
Company in FDA record
Nuvasive Specialized Orthopedics Inc
FDA record creation date
Oct 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During explantation, standard removal tools can generate sufficient forces to cause distal plug to disassociate from the bone transport system.

Source facts

Action
On 7/28/2020, the firm sent a "Field Safety Notice" to US and OUS Surgeons, and Distributors via email informing them of an update removal/explantation process and reminding them to follow the Instructions for Use. Customers are instructed to: -Review, complete, sign (via DocuSign) and return the Consignee Confirmation Form that is accompanying the notification. Directions are located on the form. -Make sure Instruction for Use are followed -Forward the notice to anyone in their facility that needs to be informed. Any questions or assistance that is needed, customers are instructed to contact the Recalling Firm at complaints@nuvasive.com. On 8/6/2020, the firm sent a revised/updated "URGENT FIELD SAFETY NOTICE" to US and OUS Surgeons and Distributors via email informing them of that the internal components of the Bone Transport Nail may remain in the patient, which may lead to a revision surgery. In addition, if adverse reactions or quality problems are experienced with the use of the affected product, customers should report the adverse event to FDA's MedWatch Adverse Event Reporting program online at www.fda.gov, by regular mail to: MedWatch, Food and Drug Administration, 5600 Fishers Lane, Rockville, MD 20853 or by fax at 800-332-0178.
Initiated
2020-07-24
Root cause
Device Design
Distribution
Worldwide distribution. US nationwide, Austria, Australia, Canada, Switzerland, Denmark, Spain, France, United Kingdom, Germany, UAE, Luxembourg, Israel, Netherlands.
Recall number
Z-0069-2021
Recall status
Open, Classified
CDRH record ID
182846
FDA record created
2020-10-07
Recall event number
86150

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗