Skip to content

Public FDA record

Z-0069-2022: CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)0086000084612

Recall record Z-0069-2022 for CardioQuip, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0069-2022
Company in FDA record
CardioQuip, LLC
FDA record creation date
Oct 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery which could result in patient infection.

Source facts

Action
On 7/30/21, CardioQuip, issued a "Urgent Medical Device Correction" notification to affected consignees. The notification was provided online through CardioQuip's website and mailed letters. In addition to notifying consignees about the recall, the firm requested consignees take the following actions to reduce the risk of patient infection: 1. The optional Airflow Redirection Hood (MCH-10ARH) may be installed onto the Refrigeration Module(MCH-10RMS) as a potential mitigant to aerosolization. This device is attached overtop the Refrigeration Module and directs exhaust downwards, thereby redirecting potential aerosols downwards and away from the surgical field. Refer to the MCH Operator/Service Manual R3 2021 for installation instructions. Contact your local CardioQuip representative for assistance purchasing the optional Airflow Redirection Hood for your Refrigeration Module. 2.Ensure all modules and accessories, including the Airflow Redirection Hood and Refrigeration Module,are cleaned and disinfected at the same time as the MCH unit according to the Maintenance section on pg. 47 of the MCH Operator/Service Manual R3 2021. 3.Do NOT move modules or accessories between cooler-heater devices to prevent possible cross-contamination. 4.Please forward this notice to all device users and other parties within and outside your organizationthat need to be aware of this correction. 5. CardioQuip continues to develop a more comprehensive understanding of these risks and determine further action. If you have any questions or concerns regarding the Refrigeration Module (MCH-10RMS), optional Airflow Redirection Hood (MCH-10ARH), or proper device maintenance procedures, please contact CardioQuip via phone or email at: CardioQuip Customer Service: +1 (979) 691-0202 Monday Friday 8 A.M. 5 P.M. CT service@cardioquip.com 6. Receipt and understanding of the corrective action outlined in this letter is required from each CardioQuip customer. Please complete and return the e
Initiated
2021-06-24
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution including in the states of TX, PA, FL, CA, HI, SC, OH, AK, NY, IL, AR, WI, MI, AL, KS, NJ, GA, AZ, KY, MS, TN, LA, DE, MA, VA, CO, NE, NC, NH, ME, WV, PA, MO, MN, IN, CT, NH, NV, UT and the country of Venezuela.
Recall number
Z-0069-2022
Recall status
Open, Classified
CDRH record ID
189103
FDA record created
2021-10-07
Recall event number
88621

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.