Skip to content

Public FDA record

Z-0069-2024: Zimmer Dermatome Blades- 10-pack boxes of blades. Intended for use with the dermatome handpiece which is

Recall record Z-0069-2024 for Zimmer Surgical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0069-2024
Company in FDA record
Zimmer Surgical Inc
FDA classification date
Oct 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Skin grafts thin and non-uniform when using the affected blades. The issue would be identified at the time of use and may result in the need for additional harvests to adequately cover the area

Source facts

Action
Zimmer issued Urgent Medical Device Recall Letter (FA 2023-00208 ) to Distributors and Risk Managers and Recall FAQs to affected consignees on September 5, 2023. Letter states reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, immediately locate and quarantine affected product in your inventory. a. Your Zimmer Biomet sales representative may remove and return the affected product from your facility on your behalf. b. Alternatively, you may directly return all affected product from your facility. i. Complete Attachment 1 Certificate of Acknowledgement for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. ii. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. iii. Mark RECALL on the outside of the returned cartons 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. Upon receipt of affected product, Zimmer Biomet will credit your account. Please return a copy of the completed response form along with your returned product to ensure proper credit. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimm
Initiated
2023-09-05
Root cause
Process control
Distribution
US Nationwide Foreign: AUSTRALIA BRAZIL CHINA HONG KONG INDIA JAPAN MALAYSIA MEXICO NETHERLANDS ON SEOUL SINGAPORE TAIWAN THAILAND
Recall number
Z-0069-2024
Recall status
Open, Classified
CDRH record ID
203293
FDA classified
2023-10-06
Recall event number
93021

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗