Open original FDA record ↗Recall record
- FDA identifier
- Z-0070-04
- Company in FDA record
- Misonix, Inc
- FDA classification date
- Oct 30, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm received complaints of tingling or static shock sensation to end users caused by forcing the output connector into its receptacle misaligned. The misalignment causes additional voltage to appear on the patient probe.
Source facts
- Action
- Misonix, Inc. notified the sole direct consignee, Byron Medical Inc., of the "Corrective Action" by letter on 8/13/03. On 9/17/03, Byron Medical Inc. notified the end users by letter, and provided instructions for proper assembly and warning labels to be affixed to the units.
- Initiated
- 2003-08-13
- Root cause
- Other
- Terminated
- 2005-02-07
- Distribution
- 1 distributor in Tucson, AZ, for further distribution to end users (physicians and medical facilities) nationwide & Puerto Rico, and in the following countries: Canada, Mexico, Australia, Switzerland, Israel, Saudi Arabia, Cyprus, Greece, Turkey, Germany, Dominican Republic, Sweden, Netherlands, China, Taiwan ROC, and S. Korea.
- Recall number
- Z-0070-04
- Recall status
- Terminated
- CDRH record ID
- 29977
- FDA classified
- 2003-10-30
- Recall event number
- 27501