O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images
Source facts
Action
Medtronic Navigation contacted customers via telephone beginning September 23, 2016. Customers were informed of the affected product, problem and actions to be taken. Medtronic requests that users immediately discontinue the use of the O-Arm O2 Surgical Imaging system for automatic registration in Stealth Station navigated surgical procedures until further notice.
Medtronic Service personnel have been scheduled to begin the necessary inspections and re-calibration activities on customer systems beginning on September 23, 2016.
Initiated
2016-09-22
Root cause
Software design
Terminated
2017-05-12
Distribution
Worldwide Distribution - US Nationwide in the states of CO, KA, NE, MI, MO, OH, WA and country of Switzerland
Recall number
Z-0070-2017
Recall status
Terminated
CDRH record ID
149921
FDA classified
2016-10-11
Recall event number
75279