A custom implant manufactured entirely from titanium for a patient with a nickel allergy was requested by a surgeon. After the device was manufactured and used in the planned surgery, it was discovered internally that the implant assembly had a CoCrMo screw included as a component of the implant, which contains nickel.
Source facts
Action
The firm contacted the surgeon who received / implanted the device. The product has been implanted, so it is not subject to being returned.
Initiated
2017-04-25
Root cause
Nonconforming Material/Component
Terminated
2018-02-06
Distribution
This was a custom implant, there was only one (1) unit manufactured and distributed. The product has been implanted, so it is not subject to being returned.
Recall number
Z-0070-2018
Recall status
Terminated
CDRH record ID
159084
FDA record created
2017-11-08
Recall event number
78299