Z-0070-2019: EEA Circular Stapler with Tri-Staple Technology, Part No. TRIEEA28MT TRIEEA28XT TRIEEA31MT TRIEEA31XT Pr

Recall record Z-0070-2019 for COVIDIEN LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0070-2019
Company in FDA record
COVIDIEN LLC
FDA record creation date
Oct 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete staple formation and/or the inability to remove the device from tissue following application potentially leading to bleeding, anastomotic leak or tissue trauma. There have been no reports of serious injury related to this issue.

Source facts

Action
All U.S. consignees were notified via Federal Express or certified mail on August 17, 2018. The letter informs customers of the potential for use of a device to have an incorrect tissue gap. Required Actions: 1. Please quarantine and discontinue use of the affected item code and lots listed above. 2. Please return affected product. All unused products from the affected item codes and lots must be returned. If you purchased directly from Medtronic, return all inventory of the affected lots to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. If you purchased from a distributor, complete all fields on the form and contact your distributor directly to arrange for return of product. 3. If you have distributed EEA" staplers with Tri-Staple" technology listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Initiated
2018-08-17
Root cause
Process design
Terminated
2020-04-10
Distribution
Worldwide Distribution - US Nationwide in the states of CA, FL, IL, NC, NY, UT. and the countries of Austria, Belgium, Denmark, Finland, France, Germany, Iceland, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Reunion, Spain, Sweden, Switzerland, and United Kingdom.
Recall number
Z-0070-2019
Recall status
Terminated
CDRH record ID
167339
FDA record created
2018-10-04
Recall event number
81043

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗