Z-0070-2026: Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Numbe

Recall record Z-0070-2026 for B BRAUN MEDICAL INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0070-2026
Company in FDA record
B BRAUN MEDICAL INC
FDA classification date
Oct 22, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.

Source facts

Action
On September 29, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters dated September 24, 2025, were sent to customers. Actions to be taken: 1. Ensure full adherence to the product Instructions for Use (IFU). In particular, the instructions for secure connection of patient connectors should be followed to reduce the probability of air ingress or leakage. 2. Increase visual inspection of the bloodlines during treatment. If air ingress or leakage is observed, re-secure or tighten connection using the product IFU. In some instances, the dialysis treatment should be stopped. 3. If re-securement or tightening connections does not resolve the concern, replace bloodline at earliest clinically feasible time. 4. If possible, avoid high flow during treatment if microbubbles are present. B. Braun Medical Inc. (BBMI) Customer/User: 1. Review this notice in its entirety. Ensure that all users in your organization of the above-mentioned products, and other concerned personnel are informed about this voluntary correction. Post this notification where the affected products are stored. Note: Product is not being returned as part of this correction notification. a. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The correction is to be extended to the consumer level. 2. Return the completed Urgent Medical Device Correction Acknowledgement Form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to recalls@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). If you have any questions, contact Director, Corporate Communications at 484-240-8373. ****Update 03/24/2026****On March 5, 2026, URGENT MEDICAL DEVICE CORRECTION letters were sent update letters out to notify of additional 15 lots impacted by the recall.
Initiated
2025-09-30
Root cause
Process control
Distribution
Domestic: US Nationwide Distribution.
Recall number
Z-0070-2026
Recall status
Open, Classified
CDRH record ID
216104
FDA classified
2025-10-22
Recall event number
97609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗