Open original FDA record ↗Recall record
- FDA identifier
- Z-0071-2008
- Company in FDA record
- Neomedix Corp
- FDA classification date
- Sep 29, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Failure of irrigation flow: The pinch valve in a few pumps may not consistently open to allow irrigation flow. In over 450 surgeries, the firm confirmed one instance of a valve that did not open. The electrically activated pinch valve in the NeoMedix Trabectome I/A pump activates irrigation flow to the eye during Trabectome surgery.
Source facts
- Action
- On May 31, 2007 NeoMedix mailed a Customer letter to all consignees informing them of the hazard. The letter also indicated to the customers they NeoMedix has developed a minor modification/upgrade that is designed to assure that the pinch valve opens at the appropriate time. The customers were instructed this modification requires return of their existing pump for modification and testing at the firm's facility before its return. In the event that surgeries have already been scheduled, NeoMedix will provide a loaner unit until the suspect unit can be modified and returned to you. The letter also indicated the cost of the modification and/or use of a loaner unit will be entirely at NeoMedix expense. You will be contacted to establish the details of implementing this modification.
- Initiated
- 2007-05-28
- Root cause
- Nonconforming Material/Component
- Terminated
- 2008-09-29
- Distribution
- Nationwide to: CA, HI, IA, NY & PA
- Recall number
- Z-0071-2008
- Recall status
- Terminated
- CDRH record ID
- 72004
- FDA classified
- 2008-09-29
- Recall event number
- 48721