Backplate reservoir holders were manufactured with an incorrect configuration of connector supports. The devices were manufactured and configured with two 3/8 inch and one inch connector supports instead of one 3/8 inch and two inch connector supports (See location of supports as # 7 and # 8 on backplate attachment). The backplates were manufactured as described in manufacturing instructions a
Source facts
Action
The firm, Terumo, interviewed the Sr. Product Manager from the Marketing Department of the company on August 29, 2012. The interview included the description of the product, problem and action taken. Replacement connectors were provided to users and all units were corrected.
If you have any questions, call 734-663-4145 or Manager, Regulatory Affairs at 734-741-5576.
Initiated
2009-10-02
Root cause
Labeling design
Terminated
2013-01-29
Distribution
Worldwide distribution: USA (nationwide) including states of: CA, MI, NC, NJ, PA, and WI;and country of: EUROPE.
Recall number
Z-0071-2013
Recall status
Terminated
CDRH record ID
113066
FDA classified
2012-10-16