Z-0071-2020: BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 651165, when using BD Trucount Tubes

Recall record Z-0071-2020 for Becton, Dickinson and Company, BD Biosciences. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0071-2020
Company in FDA record
Becton, Dickinson and Company, BD Biosciences
FDA record creation date
Oct 9, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.

Source facts

Action
On 08/19/19, the recalling firm started calling and distributing Urgent Medical Device Recall notices to customers. Laboratory managers were asked to do the following: 1) Run the Abort Count Quantification Protocol. 2) Record the percentage of aborted events on the attached Customer Response Form and submit via fax, email, or online to the recalling firm regardless of the result. 3) For percent aborted events less-than 1.0 percent: your instrument is performing as expected and no further action is required once the percent aborted events has been provided to via Customer Response Form. 4) For percent aborted events greater-than-or-equal-to 1.0 percent: suspend all patient testing. A Technical Representative will contact you within 24 hours (Mon-Fri) of receiving the data (response form) to schedule a service visit. 5) Refer to the Frequently Asked Questions document for more information. Recalling firm is investigating the root cause and to identify actions to prevent recurrence. Customers with additional questions are encouraged to call: Customer/Technical Support at 844-918-0554.
Initiated
2019-08-19
Root cause
Under Investigation by firm
Terminated
2024-04-12
Distribution
U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
Recall number
Z-0071-2020
Recall status
Terminated
CDRH record ID
176095
FDA record created
2019-10-09
Recall event number
83603

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗