Open original FDA record ↗Recall record
- FDA identifier
- Z-0071-2024
- Company in FDA record
- DKK Dai-Ichi Shomei Co., Ltd.
- FDA classification date
- Oct 6, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
Source facts
- Action
- An URGENT RECALL notification letter dated 8/23/23 was sent to the customer. Actions to be Taken by the Manufacturer We will provide an additional metal cover for the LFS bracket so that even if the shaft shears off, it will not pop out from the cover as well as instructions to properly install the cover. Also, the additional metal cover provides additional support to the LFS bracket and prevents the stopper block from broken or missing even if the excessive force is added to the bracket. The physical correction will be announced later. Other Information Contact information for questions: TEL: +81-3-3969-9221 FAX: +81-3-3969-9269 Email: info@hospilite.co.jp Monday through Friday, 8:00 AM to 5:00 PM, Japan Standard Time.
- Initiated
- 2023-08-23
- Root cause
- Device Design
- Distribution
- US States: MI
- Recall number
- Z-0071-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203249
- FDA classified
- 2023-10-06
- Recall event number
- 92999