Skip to content

Public FDA record

Z-0071-2024: LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series,

Recall record Z-0071-2024 for DKK Dai-Ichi Shomei Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0071-2024
Company in FDA record
DKK Dai-Ichi Shomei Co., Ltd.
FDA classification date
Oct 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.

Source facts

Action
An URGENT RECALL notification letter dated 8/23/23 was sent to the customer. Actions to be Taken by the Manufacturer We will provide an additional metal cover for the LFS bracket so that even if the shaft shears off, it will not pop out from the cover as well as instructions to properly install the cover. Also, the additional metal cover provides additional support to the LFS bracket and prevents the stopper block from broken or missing even if the excessive force is added to the bracket. The physical correction will be announced later. Other Information Contact information for questions: TEL: +81-3-3969-9221 FAX: +81-3-3969-9269 Email: info@hospilite.co.jp Monday through Friday, 8:00 AM to 5:00 PM, Japan Standard Time.
Initiated
2023-08-23
Root cause
Device Design
Distribution
US States: MI
Recall number
Z-0071-2024
Recall status
Open, Classified
CDRH record ID
203249
FDA classified
2023-10-06
Recall event number
92999

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗