Stryker Orthopaedics had identified that specific lots of Trident PSL Acetabular shells may have a dimensional discrepancy. The deviation regarding the difference in wall thickness will increase the gap between the shell and liner on one side and will decrease the gap between shell and liner on the opposing side, resulting in interference.
Source facts
Action
Recall notification letters were sent out on August 30, 2007 with return receipt.
Initiated
2007-08-30
Root cause
Process control
Terminated
2008-09-05
Distribution
Product was distributed nationwide and one unit to Brazil.
Recall number
Z-0072-2008
Recall status
Terminated
CDRH record ID
64356
FDA classified
2007-10-25
Recall event number
44780