Open original FDA record ↗Recall record
- FDA identifier
- Z-0072-2012
- Company in FDA record
- Cardiocommand Inc.
- FDA classification date
- Oct 19, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
CardioCommand, Inc. is recalling their TAPSCOPE 550 esophageal pacing/recording catheters because there maybe electrical discontinuity, which can result in the inability to reliably pace and/or record ECG.
Source facts
- Action
- On October 25, 2010, an Urgent Medical Device Recall Notice was sent by certified mail to all facilities advising them of the recall. The letter identified the affected product and reason for the recall. Users were asked to discontinue use and return the affected product to the firm. If all product was already used, then this should be indicated on the form provided. The firm will send replacement product in quantities equal to the returned product. The attached form should be completed and returned via fax. Customers were instructed to report any suspected incidents related to this issue to CardioCommand Customer Service at 1-800-231-6370. Also questions or comments should be directed to CardioCommand Customer Service.
- Initiated
- 2010-10-25
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-08-28
- Distribution
- Nationwide Distribution -- GA, OR, TX, CO, KY, MT, MA, AL, CA, FL & TN.
- Recall number
- Z-0072-2012
- Recall status
- Terminated
- CDRH record ID
- 100501
- FDA classified
- 2011-10-19
- Recall event number
- 58302