Ameditech is recalling the ArcPoint Labs Dip Drug Screen Dip Card because the positive interpretation illustration was incorrect on the cap labels as well as on the card cap graphic in the package insert.
Source facts
Action
A customer notification letter dated 9/23/14 was sent to all customers who purchased the ArcPoint Labs Dip Drug Screen Dip Card. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the Verification Form. Customers with questions are instructed to contact Ameditech at (858) 805-8714 or email at amd.complaint@alere.com.
Initiated
2014-09-23
Root cause
Labeling Change Control
Terminated
2015-02-27
Distribution
Distributed in AL.
Recall number
Z-0072-2015
Recall status
Terminated
CDRH record ID
130241
FDA classified
2014-10-16
Recall event number
69333