Z-0073-2018: Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model 9-201-00 Autolith

Recall record Z-0073-2018 for Northgate Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0073-2018
Company in FDA record
Northgate Technologies, Inc.
FDA record creation date
Nov 8, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The inner and outer labels on the product contain incorrect reference number of 9-202-3751DS. According to Northgate Technologies, the resistor in the probe will not allow the mislabeled product to be used with the incorrect generator. Could Delay Care.

Source facts

Action
Northgate Technologies sent an Urgent Medical Device Recall Notification dated May 24, 2017. Instructions included to examine records and locate affected devices, cease distribution or use and quarantine any affected products, return any affected product to Northgate, and contact customers if the affected product has been further distributed. For further questions, please call (800) 348-0424.
Initiated
2017-05-24
Root cause
Labeling mix-ups
Terminated
2020-06-09
Distribution
US Distribution to the following states : MA, NE and TN and Internationally to Canada.
Recall number
Z-0073-2018
Recall status
Terminated
CDRH record ID
155847
FDA record created
2017-11-08
Recall event number
77367

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗