Z-0075-2024: Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery.

Recall record Z-0075-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0075-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Oct 18, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from use of the UHI-4 during the procedures.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTIVE ACTION notification letter dated 9/22/23 was sent to customers. Action steps to be taken by the end user: Our records indicate that your facility has purchased one or more of the Olympus UHI-4. Olympus requires you to take the following actions: 1. Carefully read the content of this Field Corrective Action Letter. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this letter. 3. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at olympus5578@sedgwick.com or by fax at 866-367-6717. For any questions about the acknowledgement form, please call the Sedgwick team at 855-215-4972. 4. If you have further distributed this product, identify your customers, and forward them this Letter. Please appropriately document your notification process and let us know the end-customer feedback accordingly. Olympus requests that you report complaints, including any injuries associated with over insufflation during the procedure with UHI-4, to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. An URGENT: MEDICAL DEVICE CORRECTIVE ACTION - UPDATE notification letter dated 10/25/23 was sent to customers. Action steps to be taken by the end user: Our records indicate that your facility has purchased one or more of the Olympus UHI-4. Olympus requires you to take the following actions: 1. Inspect your inventory and identify any UHI-4 High flow insufflation unit(s). Quarantine and mark your units appropriately by your site to prevent usage. If your facility does not have alternatives or is unable to obtain alternatives, you may choose to use the UHI-4 while using extreme caution and after reviewing the information above
Initiated
2023-09-22
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-0075-2024
Recall status
Open, Classified
CDRH record ID
202943
FDA classified
2023-10-18
Recall event number
92932

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗