Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Source facts
Action
Firm began notifying customers on August 31, 2026 via URGENT MEDICAL DEVICE RECALL letters.
Customers were instructed to destroy any affected product on hand. The firm will replace any product still within its shelf life.
Initiated
2026-08-31
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-0075-2027
Recall status
Open, Classified
CDRH record ID
222243
FDA classified
2026-10-02
Recall event number
99704