Open original FDA record ↗Recall record
- FDA identifier
- Z-0076-2011
- Company in FDA record
- Impac Medical Systems Inc
- FDA classification date
- Oct 18, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software issue-- matching criteria can be configured that permits an invalid patient match, resulting in overwriting of patient records with information from a different patient.
Source facts
- Action
- Elekta Impac Software issued a Important Safety Notice letter dated September 2010 to all customers identifying the software issue and clinical impact. Customer Support representatives will contact each site and modify site configuration or upgrade existing software to the version containing the corrective fix. Customers can contact Impac Software at 800 488-4672
- Initiated
- 2010-09-02
- Root cause
- Software design
- Terminated
- 2011-06-24
- Distribution
- Worldwide Distribution: throughout the US and the countries of Australia, Austria, Belgium, Canada, China, Hong Kong, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Jordan, Morocco, Netherlands, New Zealand, Poland, Portugal, Singapore, Spain, Sweden, Switzerland, and UK.
- Recall number
- Z-0076-2011
- Recall status
- Terminated
- CDRH record ID
- 94410
- FDA classified
- 2010-10-18
- Recall event number
- 56736