Skip to content

Public FDA record

Z-0076-2012: MICROPLASTY TAPERLOC 7.5 MM RESECTION GUIDE, STAINLESS STEEL, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE PO BO

Recall record Z-0076-2012 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0076-2012
Company in FDA record
Biomet, Inc.
FDA classification date
Oct 20, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Biomet has initiated this action following an investigation which identified that the blueprint calls for engraved measurements on both sides on the Microplasty Taperloc 7.5MM Resection Guide. The above lots, as supplied to Biomet by a vendor, are engraved on only one side. To date, Biomet has received 2 reports involving 2 units where the instrument was missing the engraving on one side.

Source facts

Action
Biomet sent an "URGENT MEDICAL DEVICE RECALL NOTICE" dated September 16, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to locate and discontinue use of the affected product. Arrangements will be made for customers to return the affected product to Biomet. To confirm receipt of this notice call Biomet at 800-348-9500 ext. 3755 or 3756. Questions related to this notice should be directed to 574-371-3755 or 3756, Monday through Friday, 8 AM to 5 PM.
Initiated
2011-09-16
Root cause
Nonconforming Material/Component
Terminated
2013-03-07
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL, CA, IN, MI, OK, and VA and countries of Australia and Chile.
Recall number
Z-0076-2012
Recall status
Terminated
CDRH record ID
103982
FDA classified
2011-10-20
Recall event number
59911

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗