Z-0076-2022: Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-

Recall record Z-0076-2022 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0076-2022
Company in FDA record
Datascope Corp.
FDA classification date
Oct 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.

Source facts

Action
On Wednesday, September 8, 2021, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that all lots of Maquet Cardiosave Hybrid and Maquet Cardiosave Rescue experience the following issue: Helium Indicator Inaccuracy in the Presence of Continuous EMC Disturbances. Customers were provided with actions to take to limit the possibility of loss of therapy due to an over-reporting of remaining helium on the Helium Indicator. Cardiosave Hybrid users should view the mechanical gauge on the Cardiosave Top Panel to assess remaining helium capacity in case of suspected EMC disturbances. Cardiosave Rescue users should ensure the internal helium reservoir is full prior to use of the device. If the on-screen helium indicator is not showing full, fill the internal helium reservoir via the hospital cart or Helium Refilling Station. Please review the Cardiosave IABP Operation Transport Quick Reference Guide (0002-08-9772) for details on using the Cardiosave Rescue in transport. For both devices: Where practical maximize separation of Cardiosave system from other electronic equipment, communication devices and cables (e.g. power and communication cables). In particular, follow the recommended separation distances between communications equipment and the Cardiosave equipment as specified in the Cardiosave Operating instructions manual. The firm is developing a hardware correction to address this issue. A Datascope/Getinge service representative will contact customers to schedule the installation of the correction when available. The correction approximately will begin late Q2 2022. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2021-09-08
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution.
Recall number
Z-0076-2022
Recall status
Open, Classified
CDRH record ID
189166
FDA classified
2021-10-07
Recall event number
88651

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗