Z-0076-2027: Brand Name: ADAPTABLE Beach Chair Product Name: Beach Chair surgical patient positioner Model/Catalog N

Recall record Z-0076-2027 for QUANTUM OPS Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0076-2027
Company in FDA record
QUANTUM OPS Inc
FDA classification date
Oct 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue (inadequate weld) for patient head positioning system.

Source facts

Action
On 07/24 and 07/27/2026, an "URGENT: Medical Device Recall" letter was emailed to customers informing them that, the firm has received complaints describing the fracture of a welded joint below the head rest which provides support to patient s head. Customers are instructed to: 1. Inform individuals within organization who need to be aware, and pass this notice on to any organization where the product has been transferred. 2. Check internal inventory to locate the affected product. 3. Perform the device safety check as detailed in Attachment B and complete the Acknowledgement and Receipt Form (Attachment A) and return to SurgicalFieldActions@enovis.com or endorecall@stryker.com 4. If the device passes, it is suitable to continue clinical use. It will be scheduled for exchange with an upgraded unit. a. Upon completing the Acknowledgement and Receipt Form (Attachment A), contact Enovis Customer Service at 800-520-8976 or email SurgicalFieldActions@enovis.com to schedule the exchange, or b. Upon completing the Business Reply Form (Attachment A), Stryker will contact the account to schedule the exchange. 5. If the device fails, segregate the device and do not continue use. a. Upon completing the Acknowledgement and Receipt Form (Attachment A), please contact Enovis Customer Service at 800-520-8976 or email SurgicalFieldActions@enovis.com to arrange for return and issuance of replacement (upon availability), or b. Upon completing the Business Reply Form (Attachment A), Stryker will contact your account to arrange for return and issuance of replacement (upon availability). -The returned device will be replaced with an upgraded unit. 6. If affected product is NOT found: a. Please complete Acknowledgement and Receipt Form (Attachment A). 8. Maintain awareness of this communication internally until all required actions have been completed within their facility. Response is required.
Initiated
2026-07-24
Root cause
Device Design
Distribution
U.S. Nationwide distribution in the states of CA and TX.
Recall number
Z-0076-2027
Recall status
Open, Classified
CDRH record ID
221450
FDA classified
2026-10-02
Recall event number
99384

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗