Z-0077-2011: DDC-6-AA Data Card, an accessory to the Sentry DDU-1 00 semiautomatic external defibrillator (AED). The D

Recall record Z-0077-2011 for Defibtech, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0077-2011
Company in FDA record
Defibtech, LLC
FDA classification date
Oct 18, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Data cards may include an incorrect format which may cause the AED to not function correctly.

Source facts

Action
Defibtech Initially notified customers by telephone on 8/27/10. Customers were notified of the product, the problem and the action to be taken. Customers were instructed to locate affected data cards and send them back to Defibtech for evaluation. If the data card was inserted into the AED, then the customer was instructed to check the AED software version. If the software version was at 2.005, the customer was instructed to keep the AED out of service until Defibtech could evaluate the returned data card. (Any other software version indicated that the device was not affected by a faulty data card). If the software version was any other version other than 2.005, then the customer was instructed to keep the AED in service unless the original service inquiry required it to remain out of service. If the data card was not inserted into the AED, then the customer was instructed to return the data card for evaluation. For any questions customers were to contact Customer Service at (866) 333-4241 from 8:30-5pm EST, M-F.
Initiated
2010-08-27
Root cause
Mixed-up of materials/components
Terminated
2011-10-05
Distribution
USA - Nationwide Distribution and Puerto Rico
Recall number
Z-0077-2011
Recall status
Terminated
CDRH record ID
94386
FDA classified
2010-10-18
Recall event number
56729

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗