Open original FDA record ↗Recall record
- FDA identifier
- Z-0077-2012
- Company in FDA record
- Medtronic Perfusion Systems
- FDA classification date
- Oct 20, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The DFU for the Medtronic Pressure Display Box, model 66000, states that, for display readings of -100 to +300 mmHg there will be an accuracy of +/-5 mmHg, and for display readings of +301 to +500 mmHg there will be an accuracy of +/-10 mmHg. Through reports from customers, Medtronic has become aware that the Pressure Display Box may not always accurately display readings. During use in the -40
Source facts
- Action
- Medtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.
- Initiated
- 2011-09-09
- Root cause
- Device Design
- Terminated
- 2012-10-11
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and Yemen.
- Recall number
- Z-0077-2012
- Recall status
- Terminated
- CDRH record ID
- 104063
- FDA classified
- 2011-10-20
- Recall event number
- 59948