Open original FDA record ↗Recall record
- FDA identifier
- Z-0077-2019
- Company in FDA record
- Instrumentation Laboratory Co.
- FDA record creation date
- Oct 5, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for increased imprecision, out of range quality controls and prolonged sample results.
Source facts
- Action
- On August 1, 2018, Instrumentation Laboratory issued Urgent Medical Device Removal and Field Safety notices to their customers. Customers were advised to take the following actions: - Run quality controls with each vial and a minimum of every 8 hours per labeled insert instructions to identify possible vials with the above performance issues until an alternative product lot of HemosIL ReadiPlasTin or an alternative PT reagent has been received and is ready for use. NOTE: ACL TOP Family and ACL TOP Family 50 Series analyzers can be configured to automatically perform QC at vial change. Reference "Before vial use" under the QC Setup Definition section of the On-line Help or contact your IL representative for assistance. - Discard any vial with failed quality controls. Only use vials where all quality controls are in range. - Check your inventory for Lot No. N0278177 and add instructions (copy of the notice) to run quality controls with each vial. - Contact your local representative to convert to an alternative product lot of HemosIL ReadiPlasTin or an alternative PT reagent. - Verify the alternative lot immediately on receipt and then destroy any unused boxes of Lot No. N0278177. - Document the destruction on the Mandatory Tracking Response Record and return the completed and signed form to IL Regulatory either via fax to 781-861-4207 or via email to ra-usa@ilww.com. - Share the information with all laboratory staff and follow your internal procedures. - Forward the notification to all affected locations within your facility. - Retain a copy of the notification for your records.
- Initiated
- 2018-08-01
- Root cause
- Material/Component Contamination
- Terminated
- 2019-04-16
- Distribution
- US Nationwide
- Recall number
- Z-0077-2019
- Recall status
- Terminated
- CDRH record ID
- 167053
- FDA record created
- 2018-10-05
- Recall event number
- 80932