Z-0077-2019: Hemosll ReadiPlasTin (10 mL Size), Part Number 0020301300 - Product Usage - HemosIL ReadiPlasTin is an i

Recall record Z-0077-2019 for Instrumentation Laboratory Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0077-2019
Company in FDA record
Instrumentation Laboratory Co.
FDA record creation date
Oct 5, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for increased imprecision, out of range quality controls and prolonged sample results.

Source facts

Action
On August 1, 2018, Instrumentation Laboratory issued Urgent Medical Device Removal and Field Safety notices to their customers. Customers were advised to take the following actions: - Run quality controls with each vial and a minimum of every 8 hours per labeled insert instructions to identify possible vials with the above performance issues until an alternative product lot of HemosIL ReadiPlasTin or an alternative PT reagent has been received and is ready for use. NOTE: ACL TOP Family and ACL TOP Family 50 Series analyzers can be configured to automatically perform QC at vial change. Reference "Before vial use" under the QC Setup Definition section of the On-line Help or contact your IL representative for assistance. - Discard any vial with failed quality controls. Only use vials where all quality controls are in range. - Check your inventory for Lot No. N0278177 and add instructions (copy of the notice) to run quality controls with each vial. - Contact your local representative to convert to an alternative product lot of HemosIL ReadiPlasTin or an alternative PT reagent. - Verify the alternative lot immediately on receipt and then destroy any unused boxes of Lot No. N0278177. - Document the destruction on the Mandatory Tracking Response Record and return the completed and signed form to IL Regulatory either via fax to 781-861-4207 or via email to ra-usa@ilww.com. - Share the information with all laboratory staff and follow your internal procedures. - Forward the notification to all affected locations within your facility. - Retain a copy of the notification for your records.
Initiated
2018-08-01
Root cause
Material/Component Contamination
Terminated
2019-04-16
Distribution
US Nationwide
Recall number
Z-0077-2019
Recall status
Terminated
CDRH record ID
167053
FDA record created
2018-10-05
Recall event number
80932

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗