Reason stated in the source

Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters which were not designed for use with the VC-10, which can lead to serious injuries as it does not fit properly with the VC-10.

Source facts

Action
On September 16, 2022, the recalling firm notified affected customers through "URGENT Medical Device Field Corrective Action" letters. Customers were informed that the following new warning will be added to the IFU: Do not use replacement parts or accessories that are not compatible with the VC-10 system. Use of noncompatible accessories could lead to reduced system functionality and/or patient injury such as perforation, bleeding, and may require additional intervention. Refer to Section 11 for all accessories compatible with the VC-10 system. Customers should notify their users about risks associated with using non-compatible collection canisters and/or accessories and remind them to follow the list of compatible accessories outlined in Section 11 of the IFU. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. " Go to https://olympusamerica.com/recall. Enter the recall number 0414 " Enter your name and mailing address. Olympus will supply a copy of the revised IFU at a later date If you have distributed these devices outside your facility, please notify your customers of this matter immediately. If you have any questions, call (647) 999-3203 or email: Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Initiated
2022-09-16
Root cause
Labeling design
Distribution
Domestic distribution nationwide. Foreign distribution to Canada and Japan.
Recall number
Z-0078-2023
Recall status
Open, Classified
CDRH record ID
196110
FDA classified
2022-10-12
Recall event number
90876

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗