Open original FDA record ↗Recall record
- FDA identifier
- Z-0078-2027
- Company in FDA record
- PHILIPS MEDICAL SYSTEMS
- FDA classification date
- Oct 2, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Source facts
- Action
- On August 21, 2026, URGENT Medical Device Correction letters were sent to customers. Customer actions: User Instructions Philips recommends the following actions be taken for continued safe use of the device: 1. Continue to use your system(s) in accordance with the intended use; 2. Circulate this notice to all users of the device so that they are aware of the product issues, associated potential hazards, and manufacturer recommendations (refer to Table 1 and Table 2 for specific details regarding recommended actions); 3. Post this notice near the affected unit(s) for ease of reference until a solution is installed on your system. 4. Please complete and return the attached response form within 30 days to confirm receipt and understanding of this notice.
- Initiated
- 2026-08-21
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MD, MN, NC, NY, OH, PA, TX, VA, WV and the countries of Australia, Austria, Belgium, Bulgaria, China, Croatia, Czech Republic, Denmark, Ecuador, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lithuania, Macao, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Nicaragua, Norway, Palestine, Panama, Philippines, Poland, Portugal, Romania, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.
- Recall number
- Z-0078-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222502
- FDA classified
- 2026-10-02
- Recall event number
- 99796