Open original FDA record ↗Recall record
- FDA identifier
- Z-0079-2019
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA record creation date
- Oct 5, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result.
Source facts
- Action
- ORTHO PLUSSM is e-Communications, a service that provides subscribing customers with Ortho Clinical Diagnostics (Ortho) product notifications electronically on 15 August 2018, a customer letter (Ref. CL2018-174) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who received VITROS 5600 Integrated Systems to inform them of the potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result. Ortho also provided specific actions for customers to take when they observe unexpected MicroSlide assay results for a sample associated with TE1-504 and/or TE1-594 condition codes. Customers were also instructed to discuss any concerns related to this issue and previously reported results with the Laboratory Medical Director.
- Initiated
- 2018-08-15
- Root cause
- Software design
- Terminated
- 2022-01-10
- Distribution
- Nationwide Foreign: France, Brazil, Chile, Great Britain, Phillippines, Spain
- Recall number
- Z-0079-2019
- Recall status
- Terminated
- CDRH record ID
- 167125
- FDA record created
- 2018-10-05
- Recall event number
- 80959