Skip to content

Public FDA record

Z-0079-2021: CERTAIN BELLATEK TITANIUM ABUTMENT 6.0MM Reference Number: IEDAT6

Recall record Z-0079-2021 for Biomet 3i, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0079-2021
Company in FDA record
Biomet 3i, LLC
FDA record creation date
Oct 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may cause misalignment, the definitive abutment will not seat in the correct orientation upon attempted placement with the patient

Source facts

Action
Zimmer Biomet issued Urgent Medical Device Recall letter dated 8/31/2020 to Dental Labs and Clinicians stating reason for recall, health risk and action to take: Quarantine any affected products (definitive BellaTek abutments/3D printed models) from Attachment 1 in your possession for return to Zimmer Biomet Dental. 2. Identify any clinicians who have received an affected BellaTek Definitive Abutment as identified in Attachment 1. 3. For clinicians that have received an affected BellaTek Definitive Abutment from you, provide a copy of this notice to them along with Attachment 3 for them to complete. B. Certificate of Acknowledgement 1. Complete Attachment 2 Certificate of Acknowledgement and email to orporateQuality.PostMarket@zimmerbiomet.com. Return of product 1. Include a copy of Attachment 2 Certificate of Acknowledgement with any product tobe returned. Using the return shipping label included with this notice, please returnaffected product to the address below. Customer Service will provide additionalshipping labels if needed. Field Action PM Regulatory Compliance Zimmer Biomet 4555 Riverside Dr. Palm Beach Gardens, FL 33410 US D. Retain a copy of the Acknowledgement Form with your recall records in the event of a compliance audit of your facilitys documentation. E. If you have any questions after reviewing this notice, please call or email the Customer Service team at: 1-888-800-8045 or DentalCSDigital@ZimmerBiomet.com.
Initiated
2020-08-31
Root cause
Device Design
Terminated
2022-07-06
Distribution
FL, GA, ID, IL, MI, NV, NY, OH, SC, TN, VA
Recall number
Z-0079-2021
Recall status
Terminated
CDRH record ID
183645
FDA record created
2020-10-08
Recall event number
86402

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗