Open original FDA record ↗Recall record
- FDA identifier
- Z-0079-2023
- Company in FDA record
- Steris Corporation
- FDA classification date
- Oct 12, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The scalpel holder may separate from the handle for certain lots. If the tip falls off during use, it could fall into the patient's mouth which may result in a cut to the patient's mouth and also presents a choking hazard.
Source facts
- Action
- Starting September 2, 2022, Hu-Friedy notified all affected customers and distributors via Urgent Medical Device Recall letters. Customers were instructed to immediately destroy all on-hand inventory of the affected lots of K360 Scalpel Handles. Distributors should identify and notify any of their customers who have received further distributed product. Hu-Friedy will issue credit upon receipt of completed response forms.
- Initiated
- 2022-09-02
- Root cause
- Process design
- Terminated
- 2025-01-02
- Distribution
- Distribution to US states of IL, MA, NC, SC, and TN, Canada, China, Japan, Malaysia, Turkey, Great Britain, and India.
- Recall number
- Z-0079-2023
- Recall status
- Terminated
- CDRH record ID
- 195987
- FDA classified
- 2022-10-12
- Recall event number
- 90837