Open original FDA record ↗Recall record
- FDA identifier
- Z-0079-2027
- Company in FDA record
- Cardinal Health 200, LLC
- FDA classification date
- Oct 2, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.
Source facts
- Action
- Firm began notifying affected customers on August 19, 2026 via letters titled "Product Advisory." Customers were provided with the Medtronic/Stryker "URGENT Medical device correction UPDATE" letter and instructed to alert all affected clinicians. Customers were informed that in September 2025, Medtronic issued an update to the Instructions for Use (IFU) for the affected devices to include a relevant Adverse Events section. Customers may continue using the affected devices. Product does not need to be returned or discarded.
- Initiated
- 2026-08-19
- Root cause
- Nonconforming Material/Component
- Distribution
- Pending
- Recall number
- Z-0079-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222558
- FDA classified
- 2026-10-02
- Recall event number
- 99801