Z-0080-2019: VITROS 5600 Integrated System- (Refurbished). For use in the in vitro quantitative, semi-quantitative, an

Recall record Z-0080-2019 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0080-2019
Company in FDA record
Ortho-Clinical Diagnostics
FDA record creation date
Oct 5, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result.

Source facts

Action
ORTHO PLUSSM is e-Communications, a service that provides subscribing customers with Ortho Clinical Diagnostics (Ortho) product notifications electronically on 15 August 2018, a customer letter (Ref. CL2018-174) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who received VITROS 5600 Integrated Systems to inform them of the potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result. Ortho also provided specific actions for customers to take when they observe unexpected MicroSlide assay results for a sample associated with TE1-504 and/or TE1-594 condition codes. Customers were also instructed to discuss any concerns related to this issue and previously reported results with the Laboratory Medical Director.
Initiated
2018-08-15
Root cause
Software design
Terminated
2022-01-10
Distribution
Nationwide Foreign: France, Brazil, Chile, Great Britain, Phillippines, Spain
Recall number
Z-0080-2019
Recall status
Terminated
CDRH record ID
167126
FDA record created
2018-10-05
Recall event number
80959

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗