Open original FDA record ↗Recall record
- FDA identifier
- Z-0080-2023
- Company in FDA record
- LUMINOSTICS, INC dba CLIP HEALTH
- FDA classification date
- Oct 12, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
COVID test kit shelf life was periodically extended, as real-time stability test timepoints passed, without FDA concurrence.
Source facts
- Action
- On 09/02/2022, Field Safety notices were emailed to customers who were asked to do the following: 1) Locate affected devices in inventory and go to expiry.clipcovid.com to determine the expiration date. 2) Acknowledge receipt and understanding of the notice by emailing sales@cliphealth.com. 3) Share this notice with all those who need to be aware within your organization, or to any organization where the subject devices have been transferred. 4) If you are a distributor, please send proof to Clip Health by September 16th, of your notification to your customers, and their response. Questions or concerns can be emailed to sales@cliphealth.com On 09/21/2022, an updated field safety correction notice was sent to customers. In addition to the information above, customers were asked to do the following: a) Product beyond expiry, as shown on expiry.clipcovid.com, should be quarantined. b) Check this website periodically for potentially extended expiration dates. c) Ensure products are within expiration date claims before use. d) Acknowledge receipt and understanding of the notice using the link provided on the notice.
- Initiated
- 2022-09-02
- Root cause
- Incorrect or no expiration date
- Distribution
- US nationwide distribution
- Recall number
- Z-0080-2023
- Recall status
- Open, Classified
- CDRH record ID
- 196063
- FDA classified
- 2022-10-12
- Recall event number
- 90852