Z-0081-04: Sonotrode Probe, a straight nephroscopy probe with a 3.5 mm diameter and 362 mm working length; an access

Recall record Z-0081-04 for Richard Wolf Medical Instrument Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0081-04
Company in FDA record
Richard Wolf Medical Instrument Corp
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The sonotrode tube could become blocked by the stone fragments, causing overheating and damage to the device.

Source facts

Action
Recalled from the hospitals by sales representative telephone calls on 10/9/03 and follow-up visits to retriev the recalled lot and replace the units with a new lot of sonotrodes.
Initiated
2003-10-09
Root cause
Other
Terminated
2004-02-10
Distribution
Nebraska, Oregon, Minnesota and Illinois.
Recall number
Z-0081-04
Recall status
Terminated
CDRH record ID
29952
FDA classified
2004-07-20
Recall event number
27489

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗