Skip to content

Public FDA record

Z-0081-2014: Cordis OPTEASE Vena Cava Filter Is indicated for use in the prevention of recurrent pulmonary embolism (

Recall record Z-0081-2014 for Cordis Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0081-2014
Company in FDA record
Cordis Corporation
FDA classification date
Oct 25, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cordis identified a printing error on one unit of the Optease Vena Cava Filter, in which the orientation arrow for the femoral approach was printed in the incorrect direction. The error resulted in the filter being implanted upside down, requiring and additional percutaneous procedure to retrieve the filter.

Source facts

Action
Cordis sent an Urgent Medical Device Recall letter dated October 8, 2013, to all affected customers. The letter identified the product (Cordis OPTEASE Vena Cava Filter Catalog Numbers 466F220A and 466F220B), the problem, and the action to be taken by the customer. Customers were instructed to: 1) Read this Urgent Medical Device Recall letter. 2) Immediately identify and set aside all product listed below in a manner that ensures the affected product will not be used. 3) Review, complete, sign and return the enclosed Acknowledgement Form directly to Cordis at the fax number on the form. 4) Return any affected product with reference to the RA# on the form. Contact Customer Service (see Available Assistance section) if assistance is needed for return of the affected product. Credit will be provided or consignment levels reduced. 5) Share this letter with others in your facility that need to be made aware of this recall. 6) Contact any other facility to arrange the return of OPTEASE if any product listed below has been forwarded to them. 7) Maintain awareness of this notice until all affected product has been returned to Cordis. 8) Keep a copy of this notice with the affected product. Please contact the special Customer Service number (800) 551-7683, Monday through Friday from 8:30 AM to 6:30 PM ET. o Return instructions are provided on the Acknowledgment Form. Questions related to the Acknowledgement Form " Please contact Cordis QA at (786) 313-8730. For questions regarding this recall call 1-800-551-7683.
Initiated
2013-09-27
Root cause
Error in labeling
Terminated
2014-09-23
Distribution
Worldwide Distribution - US (nationwide) including Hawaii and Puerto Rico and states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and countries of: Canada and Panama.
Recall number
Z-0081-2014
Recall status
Terminated
CDRH record ID
122464
FDA classified
2013-10-25
Recall event number
66483

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗