Open original FDA record ↗Recall record
- FDA identifier
- Z-0082-2014
- Company in FDA record
- Maquet Medical Systems USA
- FDA classification date
- Oct 25, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.
Source facts
- Action
- MAQUET issued their customers "Urgent-Medical Device Field Correction" notices/Customer Fax Response forms dated September 16, 2013 via FedEx on September 16, 2013. The notification informs the customer of the issue with the product; how to identified affected product; and the action needed to be taken by the customer. The field correction includes service on the affected product. Authorized MAQUET Service Technicians will service the affected product to correct the issue. Customers were requested to complete and fax back (or email a scanned copy) of the enclosed response form. Customers could contact their local field representative or the MAQUET Customer Service line at 1.888.627.8383 (press option 2 followed by option 2) Monday through Friday, 8am to 6pm EST. For questions regarding this recall call 973-709-7000.
- Initiated
- 2013-09-12
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2017-06-07
- Distribution
- Nationwide Distribution including AZ, FL, CO, TX, OH, VA, MN, WI, SC, IL, ND, UT, CA, OR, NY, IA, CT, SD, MI, HI, AL, MD, KS, MO
- Recall number
- Z-0082-2014
- Recall status
- Terminated
- CDRH record ID
- 122075
- FDA classified
- 2013-10-25
- Recall event number
- 66336