Open original FDA record ↗Recall record
- FDA identifier
- Z-0082-2018
- Company in FDA record
- Invivo Corporation
- FDA record creation date
- Nov 9, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Intermittent communication between the host system and the Flex Cardio
Source facts
- Action
- On August 2017 the firm sent letters to their consignees indicating the following: This document contains important information for the continued safe and proper use of your equipment Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please retain a copy with the equipment Instruction for Use. ------------------------------------------- A problem has been detected in the Philips Xper Flex Cardio Patient Monitoring System (Flex Cardio) that, if it were to recur, could pose a risk for patients or users. This Field Safety Notice is intended to inform you about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer / user in order to prevent risks for patients or users " the actions planned by Philips to correct the problem. Because of intermittent communication between the host system and the Flex Cardio, it is possible for any of the following conditions to occur: " Inability of the Flex Cardio to connect to the host system. " Boom Monitor (display) may not display all active waveform and/or vital sign data. " Delayed audible and visual alarms at the Flex Cardio and Boom Monitor respectively, due to a mismatch between the alarm limits on the host system and the Flex Cardio/Boom Monitor. " After patient admission and case initiation at the host system, the case may not be fully transmitted to the Flex Cardio. In this case, the Boom Monitor in the exam/procedure room will display vital sign data, but the Flex Cardio will not provide audible alarms. " After patient discharge and case termination at the host system, case termination may not be transmitted to the Flex Cardio. In this case, the Boom Monitor in the exam/procedure room will continue to display vital sign data, and the Flex Cardio will continue to provide audible alarms. Our re
- Initiated
- 2017-08-11
- Root cause
- Software Design Change
- Terminated
- 2024-09-20
- Distribution
- Worldwide Distribution - US including AK, AL, AZ, CA, CO, DE, FL, GA, IA, ID, IL, KY, LA, MA, MI, MO, NC, NE, NM, NY, OH, OR, OK, PA, SD, TN, TX, VA, WA, WV and Internationally to Germany, Hungary, India, Netherlands, Russia, Saudi Arabia, South Africa and Tunisia.
- Recall number
- Z-0082-2018
- Recall status
- Terminated
- CDRH record ID
- 159623
- FDA record created
- 2017-11-09
- Recall event number
- 78465