Z-0082-2018: Xper Flex Cardio PhysioMonitoring system The Xper Flex Cardio physiomonitoring system is used to facilit

Recall record Z-0082-2018 for Invivo Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0082-2018
Company in FDA record
Invivo Corporation
FDA record creation date
Nov 9, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Intermittent communication between the host system and the Flex Cardio

Source facts

Action
On August 2017 the firm sent letters to their consignees indicating the following: This document contains important information for the continued safe and proper use of your equipment Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please retain a copy with the equipment Instruction for Use. ------------------------------------------- A problem has been detected in the Philips Xper Flex Cardio Patient Monitoring System (Flex Cardio) that, if it were to recur, could pose a risk for patients or users. This Field Safety Notice is intended to inform you about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer / user in order to prevent risks for patients or users " the actions planned by Philips to correct the problem. Because of intermittent communication between the host system and the Flex Cardio, it is possible for any of the following conditions to occur: " Inability of the Flex Cardio to connect to the host system. " Boom Monitor (display) may not display all active waveform and/or vital sign data. " Delayed audible and visual alarms at the Flex Cardio and Boom Monitor respectively, due to a mismatch between the alarm limits on the host system and the Flex Cardio/Boom Monitor. " After patient admission and case initiation at the host system, the case may not be fully transmitted to the Flex Cardio. In this case, the Boom Monitor in the exam/procedure room will display vital sign data, but the Flex Cardio will not provide audible alarms. " After patient discharge and case termination at the host system, case termination may not be transmitted to the Flex Cardio. In this case, the Boom Monitor in the exam/procedure room will continue to display vital sign data, and the Flex Cardio will continue to provide audible alarms. Our re
Initiated
2017-08-11
Root cause
Software Design Change
Terminated
2024-09-20
Distribution
Worldwide Distribution - US including AK, AL, AZ, CA, CO, DE, FL, GA, IA, ID, IL, KY, LA, MA, MI, MO, NC, NE, NM, NY, OH, OR, OK, PA, SD, TN, TX, VA, WA, WV and Internationally to Germany, Hungary, India, Netherlands, Russia, Saudi Arabia, South Africa and Tunisia.
Recall number
Z-0082-2018
Recall status
Terminated
CDRH record ID
159623
FDA record created
2017-11-09
Recall event number
78465

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗