Open original FDA record ↗Recall record
- FDA identifier
- Z-0082-2022
- Company in FDA record
- Bio-Rad Laboratories, Inc.
- FDA record creation date
- Oct 8, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.
Source facts
- Action
- On 11/01/2019, ANA Screen Lot Matching-Compatibility Notice, were mailed to customers. Customers were advised of the following: If you have received the new control material, please ensure that your laboratory is using one of the newer lot numbers of BioPlex ANA calibrators and ANA reagents. Should you have any of the older BioPlex ANA reagent packs and calibrator sets listed below in ITALICS remaining in your inventory or if you have used these lots in combination with the new control, please contact your local Technical Support center. The following telephone number was provided: 510-724-7000.
- Initiated
- 2019-11-01
- Root cause
- Process control
- Terminated
- 2023-09-27
- Distribution
- Worldwide distribution - U.S Nationwide distribution including in the states of CA, FL, NJ, IL, NE, TN, IN, TX, LA, WA, MI, MA, MN, OH, HI, MO, PA, CT, UT, SD, NY, AL, PR, OR, MD, KY, NC, AZ, SC, VA, WI, GA, IA, RI, AK, OK, KS, ND, MS, ME, NM, AR and the countries of ES, AE, DE, FR, KR, CN, TW, JP, IL, SG, AU, NZ, HK, IT, CA, PT, GB, NO, CZ, SE, CH, UA, RU, BE, DK.
- Recall number
- Z-0082-2022
- Recall status
- Terminated
- CDRH record ID
- 189173
- FDA record created
- 2021-10-08
- Recall event number
- 88652