Open original FDA record ↗Recall record
- FDA identifier
- Z-0083-2010
- Company in FDA record
- The Anspach Effort, Inc.
- FDA classification date
- Oct 28, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Anspach Effort, Inc. is recalling the Foot Control/ Foot Pedal for SC2000 and SC2100 Powered, simple cranial drills, burrs, trephines, and accessories. The Foot Control Cables lacks the shield connection from the cable to the connector body. The lack of shielding may result in radiated emissions which could possibly cause interference on other electrical equipment adjacent to the foot pedal
Source facts
- Action
- Anspach issued an "Urgent Product Removal" Letter dated August 17, 2009 informing customers of the affected product. Consignees are asked to screen their inventory and return any affected product including a completed Product Replacement Form to the firm. For further questions, contact Anspach at 1-800-327-6887 or 1-561-627-1080.
- Initiated
- 2009-08-13
- Root cause
- Device Design
- Terminated
- 2010-01-05
- Distribution
- Worldwide Distribution -- Canada, Indonesia, United Kingdom, Korea and United States (NC, OK, TN, CA, KS, MI and WA).
- Recall number
- Z-0083-2010
- Recall status
- Terminated
- CDRH record ID
- 84580
- FDA classified
- 2009-10-28
- Recall event number
- 53073