Z-0084-2022: MYLA Versions: 4.7, 4.7.1, 4.8, 4.8.1 and 4.8.2, part of the VIRTUO System

Recall record Z-0084-2022 for BioMerieux SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0084-2022
Company in FDA record
BioMerieux SA
FDA record creation date
Oct 8, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain conditions, there is a risk for a false negative result.

Source facts

Action
A customer letter was issued to impacted subsidiaries and distributors on 09/08/2021. The letter includes instructions that the customer verify that they operate under the specific conditions required for the anomaly to occur: You are using MYLA (V4.7, V4.7.1, V4.8, V4.8.1 or V4.8.2) in conjunction with BCI Connect AND You are managing Blood Culture analyses on BACT/ALERT VIRTUO connected to MYLA AND Your Blood Culture workflow is based on 2 or more bottles with the same specimen (same specimen/laboratory id) AND You are using the LIS capability to send Blood Culture requests containing the incubation expected duration (MTT) AND You are sending other requests than Blood Culture (Offline SU, ID,) on the same specimen than the one used for BC note: this condition is applicable only if it occurs after the previous ones, chronologically speaking - If the answer to ALL of the above questions is yes, then the customer must contact their bioMerieux representative for help to implement a workaround that should be used until MYLA version 4.9 is available. If the answer to any of the above question is no, the customer will not be impacted by this anomaly. The firm issued an updated Field Safety Corrective Action, FSCA #5308-2, on 9/29/2021 to subsidiaries and distributors which replaced FSCA #5308-1 to revise references in the FSCA document to only include the references of the impacted software. The firm added Annex I to the enclosed customer letter to be issued by the subsidiaries/distributors that provides guidance on how to check the MYLA software version and if BCI connectivity is being used.
Initiated
2021-09-02
Root cause
Software design
Terminated
2026-03-09
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Australia, Belgium, Canada, Switzerland, Chile, Columbia, Czech Republic, Denmark, Germany, Spain, Finland, France, United Kingdom, Guadeloupe, Hong Kong, Canary Islands, Ireland, India, Italy, Japan, South Korea, The Netherlands, Norway, Poland, Portugal, Sweden, Singapore, South Africa.
Recall number
Z-0084-2022
Recall status
Terminated
CDRH record ID
189192
FDA record created
2021-10-08
Recall event number
88658

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗