Z-0084-2027: DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520

Recall record Z-0084-2027 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0084-2027
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Oct 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

Source facts

Action
On September 2, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to inform their laboratory or medical director to assess laboratory and patient impact. For analyzers on software versions earlier than 1.5: - Users should verify that their instrument displays the intended units in the Test Menu page within the System Configuration page prior to reporting patient results. - Recall letter should be posted on or near affected analyzers if they have the affected software version. This issue is resolved in software version 1.5. If your analyzer is currently operating on an earlier software version, your Beckman Coulter service representative will contact you to schedule the upgrade to software version 1.5.
Initiated
2026-09-02
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-0084-2027
Recall status
Open, Classified
CDRH record ID
222508
FDA classified
2026-10-02
Recall event number
99800

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗