Open original FDA record ↗Recall record
- FDA identifier
- Z-0084-2027
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Oct 2, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
Source facts
- Action
- On September 2, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to inform their laboratory or medical director to assess laboratory and patient impact. For analyzers on software versions earlier than 1.5: - Users should verify that their instrument displays the intended units in the Test Menu page within the System Configuration page prior to reporting patient results. - Recall letter should be posted on or near affected analyzers if they have the affected software version. This issue is resolved in software version 1.5. If your analyzer is currently operating on an earlier software version, your Beckman Coulter service representative will contact you to schedule the upgrade to software version 1.5.
- Initiated
- 2026-09-02
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution.
- Recall number
- Z-0084-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222508
- FDA classified
- 2026-10-02
- Recall event number
- 99800